Author (up) Aura Hernandez-Sabate; David Rotger; Debora Gil edit   pdf
doi  openurl
  Title Image-based ECG sampling of IVUS sequences Type Conference Article
  Year 2008 Publication Proc. IEEE Ultrasonics Symp. IUS 2008 Abbreviated Journal  
  Volume Issue Pages 1330-1333  
  Keywords Longitudinal Motion; Image-based ECG-gating; Fourier analysis  
  Abstract Longitudinal motion artifacts in IntraVascular UltraSound (IVUS) sequences hinders a properly 3D reconstruction and vessel measurements. Most of current techniques base on the ECG signal to obtain a gated pullback without the longitudinal artifact by using a specific hardware or the ECG signal itself. The potential of IVUS images processing for phase retrieval still remains little explored. In this paper, we present a fast forward image-based algorithm to approach ECG sampling. Inspired on the fact that maximum and minimum lumen areas are related to end-systole and end-diastole, our cardiac phase retrieval is based on the analysis of tissue density of mass along the sequence. The comparison between automatic and manual phase retrieval (0.07 ± 0.07 mm. of error) encourages a deep validation contrasting with ECG signals.  
  Address Beijing (China)  
  Corporate Author Thesis  
  Publisher Place of Publication Editor  
  Language Summary Language Original Title  
  Series Editor Series Title Abbreviated Series Title  
  Series Volume Series Issue Edition  
  ISSN ISBN Medium  
  Area Expedition Conference  
  Notes IAM;MILAB Approved no  
  Call Number IAM @ iam @ HRG2008 Serial 1553  