TY - CONF AU - Aura Hernandez-Sabate AU - David Rotger AU - Debora Gil PY - 2008// TI - Image-based ECG sampling of IVUS sequences BT - Proc. IEEE Ultrasonics Symp. IUS 2008 SP - 1330 EP - 1333 KW - Longitudinal Motion KW - Image-based ECG-gating KW - Fourier analysis N2 - Longitudinal motion artifacts in IntraVascular UltraSound (IVUS) sequences hinders a properly 3D reconstruction and vessel measurements. Most of current techniques base on the ECG signal to obtain a gated pullback without the longitudinal artifact by using a specific hardware or the ECG signal itself. The potential of IVUS images processing for phase retrieval still remains little explored. In this paper, we present a fast forward image-based algorithm to approach ECG sampling. Inspired on the fact that maximum and minimum lumen areas are related to end-systole and end-diastole, our cardiac phase retrieval is based on the analysis of tissue density of mass along the sequence. The comparison between automatic and manual phase retrieval (0.07 ± 0.07 mm. of error) encourages a deep validation contrasting with ECG signals. L1 - http://refbase.cvc.uab.es/files/HRG2008.pdf UR - http://dx.doi.org/10.1109/ULTSYM.2008.0322 N1 - IAM;MILAB ID - Aura Hernandez-Sabate2008 ER -