Title | Image-based ECG sampling of IVUS sequences | Type | Conference Article | ||
Year | 2008 | Publication | Proc. IEEE Ultrasonics Symp. IUS 2008 | Abbreviated Journal | |
Volume | Issue | Pages | 1330-1333 | ||
Keywords | Longitudinal Motion; Image-based ECG-gating; Fourier analysis | ||||
Abstract | Longitudinal motion artifacts in IntraVascular UltraSound (IVUS) sequences hinders a properly 3D reconstruction and vessel measurements. Most of current techniques base on the ECG signal to obtain a gated pullback without the longitudinal artifact by using a specific hardware or the ECG signal itself. The potential of IVUS images processing for phase retrieval still remains little explored. In this paper, we present a fast forward image-based algorithm to approach ECG sampling. Inspired on the fact that maximum and minimum lumen areas are related to end-systole and end-diastole, our cardiac phase retrieval is based on the analysis of tissue density of mass along the sequence. The comparison between automatic and manual phase retrieval (0.07 ± 0.07 mm. of error) encourages a deep validation contrasting with ECG signals. | ||||